Product specifications

Model:
IRT 3030
Displayed temperature
range:
34 °C – 42.2 °C (93.2 °F – 108 °F)
Operating ambient
temperature range:
10 °C – 40 °C (50 °F – 104 °F)
Storage temperature:
-25 °C – 55 °C
Display resolution:
0.1 °C or °F
Accuracy for displayed
temperature range:
± 0.2 °C (35.5 – 42 °C)
(95.9 – 107.6 °F)
± 0.3 °C (outside this temperature range)
Clinical bias:
-0.11 °C (-0.198 °F)
Limits of agreement:
± 0.25 °C (0.45 °F)
Clinical repeatability:
± 0.10 °C (0.18 °F)
Measuring Site:
Ear
Reference Site:
Oral (estimated oral temperature is
displayed)
Operating and storage
relative humidity:
≤ 95% non-condensing
Battery life:
2 years /1000 measurements
Service life:
5 years
Warranty:
2 years
If device is not used or stored within specified temperature and
humidity ranges the technical accuracy cannot be ensured.
10
Type BF
See instructions
equipment
for use
55
25
Storage
temperature
Subject to change without notice.
IP22: Protected against solid foreign objects of 12.5 mm
diameter and greater. Protected against vertically falling water
drops when the device is tilted up to 15 degrees.
EN 12470-5:2003 Clinical thermometers—Part 5: Performance
of infra-red ear thermometers (with maximum device)
ISO 80601-2-56:2009 Medical electrical equipment—
Part 2-56: Particular requirements for basic safety and
essential performance of clinical thermometers for body
temperature measurement
This product conforms to the provisions of the EC directives
93/42/EEC (Medical Device Directive) and 2011/65/EU (RoHS).
The Braun ThermoScan needs special precautions regarding
EMC. For detailed description of EMC requirements please
refer to www.hot-europe.com/support.
Portable and mobile RF communications equipment can
affect the Braun ThermoScan.
This product contains batteries and recyclable
electronic waste. To protect the environment, do not
dispose of it in the household waste, but take it to
appropriate local collection points.
English
Operating
temperature
11

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Thermoscan irt 3030

Inhaltsverzeichnis