Herunterladen Inhalt Inhalt Diese Seite drucken

Replacing The Battery; Product Specifications - Braun ThermoScan IRT 3030 Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für ThermoScan IRT 3030:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 8

Replacing the battery

The thermometer is supplied with one 3V Lithium cell battery
(type 2032). Insert new battery when the low battery symbol
appears on the display.
The battery door design is
intended to prevent a small
child from easily opening the
battery door to address a
possible choke hazard. Open
the battery compartment by
inserting a narrow pointed object such as a ball point pen or
paperclip into the small hole above the battery door (8). While
pressing into the hole slowly slide back the battery door with
the other hand.
To close battery door align the tabs along the side of the
battery into the battery compartment and slowly slide back
into position until the door clicks into its locked position.
Calibration
This product has been calibrated at the time of manufacture.
If used according to the use instructions, periodic calibration
is not required. This device is not intended as a substitute for
regular check-ups by your doctor, please consult your doctor if
you have any doubt about the temperature reading. LOT and
Serial numbers of your product must be provided as they are
essential to record and follow up your inquiry or claim.
The LOT and Serial numbers are located in the battery
compartment.
The production date is given by the LOT number and can be
deciphered as explained below:
The first 3 numeric digits in the LOT number represent the day
of the year of manufacture. The next 2 numeric digits
represent the last two numbers of the calendar year of
manufacture and the letter(s) at the end designate the
manufacturer of the product. (E.g.: LOT No.: 12313tav this
product was made on the day 123, year 2013 at the
manufacturer code tav.)
10
Guarantee
EMC tables and Consumer Card are available on our website at
www.hot-europe.com/support.
For support requests, please visit www.hot-europe.com/support
or find service contact information on the last page of this
manual.
UK only: This guarantee in no way affects your rights under
statutory law.

Product specifications

Model:
IRT 3030
Displayed temperature
range:
34 °C – 42.2 °C (93.2 °F – 108 °F)
Operating ambient
temperature range:
10 °C – 40 °C (50 °F – 104 °F)
Storage temperature:
-25 °C – 55 °C
Display resolution:
0.1 °C or °F
Accuracy for displayed
temperature range:
± 0.2 °C (35.5 – 42 °C)
(95.9 – 107.6 °F)
± 0.3 °C (outside this temperature range)
Clinical bias:
-0.11 °C (-0.198 °F)
Limits of agreement:
± 0.25 °C (0.45 °F)
Clinical repeatability:
± 0.10 °C (0.18 °F)
Measuring Site:
Ear
Reference Site:
Oral (estimated oral temperature is
displayed)
Operating and storage
relative humidity:
≤ 95% non-condensing
Battery life:
2 years /1000 measurements
Service life:
5 years
Warranty:
2 years
If device is not used or stored within specified temperature and
humidity ranges the technical accuracy cannot be ensured.
English
11

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis