5. Packaging / 6. Starting, operation
In order to protect the diaphragm, the pressure connection of the instrument E-11 is provided
with a special plastic protection cap.
Remove this plastic protection cap only just before installing the pressure trans-
mitter in order to prevent any damage to the diaphragm or the thread.
Keep the plastic protection cap of the pressure connection thread and the
diaphragm for later storage or transport.
Mount the plastic protection cap when removing and transporting the instru-
ment.

6. Starting, operation

6.1 Diaphragm test for your safety
Required tools: wrench (flats 27), screw driver
It is necessary that before starting the pressure transmitter you test the diaphragm, as this is a
safety-relevant component.
WARNING
Pay attention to any liquid leaking out, for this points to a diaphragm damage.
Check the diaphragm visually for any damage (E-11).
Use the pressure transmitter only if the diaphragm is undamaged.
Use the pressure transmitter only if it is in a faultless condition as far as the
safety-relevant features are concerned (e.g. no damage of case, cable and
connection threads. Otherwise explosion protection is not ensured).