10. Install Disposable Cartridge in the TCU, by lining up the 2 pins on the
cartridge with the 2 slots in the TCU, and rotating it clockwise, as
shown in the figure below . Ensure that ―THIS SIDE UP‖ and the
guidance markings are visible on the top side of the catheter. Leave
the plastic balloon cover on.
WARM-UP
1. Turn the ―POWER SWITCH‖ to the ―ON‖ position; the LCD will show
the following message: If no display is present then do not continue
with procedure.
Both LED's should be OFF (see below right-bottom corner of LCD).
(If the number of treatments remaining before servicing is less than
50, the TCU will issue a beeping sound and the LCD will show the
following message for 10 seconds:
where ―X‖ is number of remaining uses of the TCU.)
2. The unit automatically performs a self-test, and, if no technical
problems are encountered, the initial message is replaced by the
following message:
(If the TCU encounters a technical problem, an error number will be
displayed on the LCD. Please refer to the Operator's Manual for
troubleshooting.)
3. The following message is then displayed:
where ―XXX‖ is the measured temperature of the treatment liquid.
The ―Heating‖ LED will turn ON, indicating that the pre-heating of the
treatment liquid has begun. The ―R eady‖ LED should be OFF. Before
entering the ―READY FOR TREATMENT‖ state the TCU performs a
leak test. If this test is not passed (―ERROR #19‖ will appear on LCD),
turn the ―POWER SWITCH‖ to the ―OFF‖ position, and verify that:
WELCOME TO
THERMABLATE
X TREATMENT(S)
BEFORE SERVICING
Device OK
HEATING – Wait...
Fluid Temp XXX °C
7