A second hysteroscopy should be performed immediately
following curettage to ensure that the curettage has not created a
perforation of the uterus.
8. Alternatively, use ultrasonic surveillance during the treatment to check
for correct balloon position inside the uterine cavity.
9. Slide the balloon cover off the balloon. DO NOT DISPOSE OF
BALLOON COVER AS IT IS REQUIRED FOR LATER USE .
Remove the Thermablate EAS system from the TCU Stand.
10. Slowly insert the Thermablate balloon until balloon tip touches the
fundus. Tap the tip of catheter gently against the fundus to confirm
placement of the catheter within the uterus.
11. Ensure that the depth marking on the balloon catheter matches the
previously obtained sounding measurements. Should there be a
discrepancy of more than 0.5 cm between the sounding
measurements obtained and depth marking on the catheter, a repeat
hysteroscopy should be performed.
12. Activate the treatment cycle by holding Blue Treatment Trigger
Switch for 5 seconds. After hearing five (5) short and one (1) long
beeping sounds, the treatment will automatically begin. The Finger
can be removed from the trigger switch at this point. The LCD will
show the following message:
Both LEDs will turn OFF.
13. After 15 seconds, if the TCU passes the system check, the actual
treatment cycle w ill start, and the LCD will show the following
message:
Followed shortly after by:
where ―X‖ is the sign (+ or -) for positive or negative pressure.
―YYY‖ is the actual pressure value reached during
―Z:ZZ‖ is the treatment time remaining (min:sec).
If th e system check fails, the TCU will issue an alarm and
stop operating. In this case, turn the "POWER SWITCH" to
the "OFF" position, d i s c o n n e c t t h e p o w e r c o r d ,
a n d remove the Disposable Cartridge from the patient,
and verify that:
Performing
System Check
TREATING...
Starting...
Pressure: XYYY
Time Left: Z:ZZ
procedure (mmHg).