A) The Disposable Cartridge has been installed properly into the
TCU, and
B) No fluid is leaking out from the Disposable Cartridge, particularly
from the balloon.
In case of a leaking Disposable Cartridge that has not spilled treatment
fluid into the TCU, replace with a new one. USE EXTRA CAUTION, AS
THE METALLIC PART OF THE DISPOSABLE CARTRIDGE MAY BE
VERY HOT.
Turn the ―POWER SWITCH‖ back to the ―ON ‖ position and follow the
instructions until reaching the system check.
If treatment fluid has leaked into the TCU, it will need to be repaired.
14. As the balloon deploys, it may push the catheter slightly backwards
(Up to 0.5 cm is normal). Do not push the catheter forward during
treatment.
At no time during the treatment should the catheter advance
beyond the pre-determined sounding length. Should this occur,
abort the procedure by activating the "Emergency Stop"button,
observe the "Treatment Stop" message, while the fluid is
actively withdrawn from the balloon and wait for the
message: "Finished V: XX ml Withdraw Balloon" to appear on
the LCD screen and slowly remove the Thermablate catheter
from the uterus. Perform a hysteroscopy to ensure that the
15. Observe the LCD screen on TCU as it automatically performs
system checks and completes the treatment cycle.
Automatic operation of the System inflates the balloon, controls
the pressure, and pulses the treatment liquid, maintaining uniform
temperature in the balloon. During this time, the pump is clearly
audible. This is not a malfunction – it is part of the normal
operation of the unit.
IN THE EVENT OF POWER LOSS DURING TREATMENT
If the power is lost during treatment, wait for thirty (30) seconds. If,
after this 30-second period, the power has not come on again,
withdraw the balloon fast but carefully since some liquid may still be
contained in the balloon. Turn the "POWER SWITCH" to the "OFF"
position, and then go to Section "Post-Treatment" below.
If the power has come on again within this 30-second period, DO
NOT WITHDRAW TH E BALLOON. The TCU will automatically
recognize that the "previous" treatment did not finish correctly and
will proceed to actively withdraw the liquid from the balloon.
REMOVE THE BALLOON FRO M THE UTERU S ONLY AFTER TH E
"FINISHED V: XX ml Withdraw Balloon" MESSAGE APPEARS
CAUTION
uterus has not been perforated.
CAUTION
ON THE LCD SCREEN.
11