Herunterladen Inhalt Inhalt Diese Seite drucken

Thermo Scientific Labofuge 200 Betriebsanleitung Seite 11

Inhaltsverzeichnis

Werbung

5.1 Usage of Commercial Sample Vessels (only for Labofuge 200)
Dimension of
sample vessels
Cat.- No.
17x100 mm
16x100 mm
16x100 mm
#3763
Vacutainer
#3761
13x75 mm
with or without
screw cap
Vacutainer
#3761
13x100 mm
with or without
screw cap
Sarstedt-
#3762
Monovettes
10 ml with
screw cap
Abb. 5/ Fig. 5
Reduzier-Adapter und 2 Distanz-Stopfen
reducing adaptor and 2 distance plugs
adaptateur de reduction et 2 insertion
a
distance
accessories
designation
purpose
protective
prevents
container
contamination
in case of
glass breakage
reducing
to hold
adaptor
short
with
13 mm tubes
distance plug
reducing
to hold
adaptor
long
without
13 mm tubes
distance plug
Monovette
for correct
adaptor
positioning of
(rubber
Monovettes
cushion)
mounting instruction
Tube can be inserted directly into the rotor borings
ATTENTION - WARNING
When the rubber cushions for Monovettes are
used, no more than 6 tubes can be inserted!
ATTENTION - WARNING!
Do not misuse the protective containers as
sample vessels!
Put the tubes into the protective containers and insert
them completely into the rotor borings
1. Disassemble the rotor (see 4.2) and remove the
Monovette adaptors if present
2. Mount the reducing adaptors with inserted
distance plug (with pins placed onto the conical
surface of the rotor lower part)
3. Re-assemble the rotor and install (see 4.3 and 4.4)
4. Insert the short 13 mm tubes through the rotor
borings into the premounted adaptors
1. Disassemble the rotor (see 4.2) and remove the
Monovette adaptors if present
2. Push out the distance plugs using a long 13 mm
tube (see Fig. 5)
3. Mount the reducing adaptors by placing the
distance pins on the conical surface of the rotor
lower part
4. Re-assemble the rotor and install (see 4.3 and 4.4)
5. Insert the long 13 mm tubes through the rotor
borings into the premounted adaptors
ATTENTION - WARNING!
Operation without Monovette adaptors provides
rotor damage!
1. Disassemble the rotor (see 4.2)
2. Press the rubber cushions with the chamfer edges
facing down into the bottom rim of the rotor lower
part
3. Re-assemble the rotor and install (see 4.3 and 4.4)
4. Load the rotor symmetrically with maximally
6 Monovette tubes