Replacement transmitter
Vorschau ausblenden Andere Handbücher für Promag 50:

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Replacement transmitter
• To regain the (explosion) protection of the measuring device:
– seal the openings with genuine covers
– seal cable entries with approved components suitable for the (explosion)
protection.
• Replace defective seal/gaskets by genuine parts from Endress+Hauser only.
• Threads (e.g. of the cover for the electronics and connection compartments)
must be lubricated. Use an acid-free, non-hardening grease if an abrasion
resistant dry lubricant is non-existent.
• If threads are damaged or defective, the measuring device must be repaired.
• If spacing is reduced or the dielectric strength of the measuring device cannot
be guaranteed during repair work, perform a test on completion of the work
(e.g. high-voltage test in accordance with the manufacturer's instructions).
• Risk of damaging electronic components!
Ensure you have a working environment protected from electrostatic
discharge.
• Danger: Risk of electric shock! Only open the measuring device
in a de-energized state.
• Observe the instructions for transporting and returning the device outlined in
the Operating Instructions.
2 Promag 10: Replacing transmitter
a
c
e
Endress+Hauser
b
A0012026
d
24
25 26
27
1
2
0.5 × 3.5
24
25 26
27
1
2
A0013088
f
T20
A0013146
g
h
3 Promag 23: Replacing transmitter
a
A0012028
c
A0013163
-
Esc
e
A0012029
S-DAT
T-DAT
Promag 10, 23, 50, 53, 55
!
12 Nm
Note!
Special screw! Do not loose screw
8 mm
completely or replace with another scew.
Only use genuine parts from
Endress+Hauser.
b
4(-) 3(+) 2(-)1(+)
4(-) 3(+) 2(-)1(+)
A0012707
d
+
E
A0013157
f
A0013160
A0013147
A0012034
0.5 × 3.5
A0013145
Ph 2
A0013145
3 mm
A0012513
2