Herunterladen Inhalt Inhalt Diese Seite drucken

Hach LDO sc Model 2 Sensor Benutzerhandbuch Seite 15

Inhaltsverzeichnis
Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 20
6.3 Exit calibration procedure
1. To exit a calibration, push back.
2. Select an option.
Option
Cancel
(or ABORT)
Return to calibration
(or BACK TO CAL)
Exit
(or LEAVE)
6.4 Set to factory calibration
To set the sensor back to the factory calibration:
1. Go to the calibration menu:
• SC4500 Controller—Select the tile of the device, then select Device menu > Calibration.
• SC200 and SC1000 Controllers—Go to the main menu, then select SENSOR SETUP > [select
instrument] > CALIBRATE.
2. Select Reset calibration (or RESET DFLT CAL).
Section 7 Sensor data and event logs
The SC Controller provides a data log and an event log for each sensor. The data log stores the
measurement data at selected intervals (user configurable). The event log shows the events that
have occurred.
The data log and event log can be saved in CSV format. Refer to the controller documentation for
instructions.
Section 8 Modbus registers
A list of Modbus registers is available for network communication. Refer to the manufacturer's
website for more information.
Section 9 Maintenance
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
Explosion hazard. Do not connect or disconnect electrical components or circuits to the equipment
unless power has been switched off or the area is known to be non-hazardous.
Explosion hazard. Substitution of components may impair suitability for Class 1, Division 2. Do not
replace any component unless power has been switched off and the area is known to be non-
hazardous.
Description
Stop the calibration. A new calibration must start from the beginning.
Return to the calibration.
Exit the calibration temporarily. Access to other menus is allowed. A calibration for a
second sensor (if present) can be started.
SC200 and SC1000 Controller—To return to the calibration, push menu and select
SENSOR SETUP > [Select Sensor].
D A N G E R
D A N G E R
D A N G E R
English 15
Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Ldo sc 2

Inhaltsverzeichnis