Resistance reading – A and B result flag (registers 506-507)
4.3
The following tables detail the result flag bytes:
bit 7 (MSB)
bit 6
0 – OK
not used
not used
1 – the
measurement
has not
begun
*
See the overview of registers 200-207, Parameter 1.
bit 7 (MSB)
bit 6
not used
not used
4.4
Status flag readout (register 520)
The bit fields of the Status byte are explained below:
bit 7 (MSB)
bit 6
not used
not used
not used
Result flag field A (register 506)
bit 5
bit 4
If the AutoRange full mode is disabled*
and the measurement function is
R
|U
= 500 V or
ISO
ISO
R
|U
= 1000 V:
ISO
ISO
0 – no underflow error (R
1 – underflow error present (R
If the AutoRange full is active*, the bit is
always set to 0.
Result flag field B (register 507)
bit 5
not used
bit 5
bit 4
0 – test
0 – no voltage ≥ 50V
specimen
detected
discharging not
1 – measurement
in progress
stopped by live voltage
1 – test
present on the R
specimen
R
discharging in
progress
The measurement is
stopped when AC ≥ 50 V
or DC ≥ 50 V with
reversed polarity is
detected at the test
terminals.
MIC-RS | MODBUS PROTOCOL – USER MANUAL
bit 3
0 – stable
output voltage
1 – unstable
output voltage
(rising)
≥ 1 kΩ)
ISO
< 1 kΩ)
ISO
bit 4
bit 3
not used
Measurement mode:
00 – normal mode
01 – single mode
10 – continuous mode
Status
bit 3
0 – no auto measure-
ment end timeout
occurred
1 – auto measurement
end timeout occurred
+,
ISO
- test terminals.
ISO
This bit is set if the
measurement is
stopped automatically
with the auto measure-
ment end timeout (see
the overview of registers
200-207).
bit 2
bit 1
0 – stable
0 – no violation
result
1 – violation
1 – unstable
present
result (see
Section 5)
bit 2
bit 1
Actual measurement function:
00 – R
X
01 – R
|U
ISO
10 – R
|U
ISO
11 – R
|U
ISO
ISO
bit 2
bit 1
0 – measure-
ments stopped,
test voltage U
ISO
off
1 – measure-
ments in progress
bit 0 (LSB)
0 – new result
1 – last result
bit 0 (LSB)
= 500 V
ISO
= 1000 V
ISO
= 10•U
V
ADJ
bit 0 (LSB)
0 – calibration
coefficients OK
1 – calibration
coefficients invalid
(bad CRC).
23