10 Cleaning and Sanitization

Caution!
Ports on the back panel are not tight.
The Sterisart Universal pump is not suited for sanitization (gaseous or atomized)
with peracetic acid, hydrogen peroxide, or other decontamination procedures
commonly used in isolators, unless it is built into an isolator countertop.
These types of disinfectant can be used when the pump has been professionally
installed in the isolator.
Make sure that no liquids penetrate the socket of the external display and
control unit. If the external display and control unit is unplugged during the
cleaning process, the supplied protection cover must be inserted into the socket.
– Before cleaning the pump rotor, turn off the main power switch and unplug
the power cable.
The following parts of the pump have to be removed for cleaning:
– Bow, lever, support rod and draining container
– Disconnect the pump head:
– The rotor must be turned by hand until the marking on the upper side (black dot)
is pointing backwards.
– The marking indicates the position of the fastening screw that is located in
the rotor mid-section.
– Loosen this screw with a 4-mm Allen wrench. Do not turn the screw out
completely.
– Pull the rotor upwards.
– Pull the speed-selecting knob upwards.
General Instructions:
Comply with the IP protection rating! No liquid may enter the Sterisart
Universal pump. Disconnect the pump from line voltage before any cleaning,
maintenance, or repair work.
Cleaning stainless steel surfaces:
The stainless steel housing can be cleaned at regular intervals. Clean stainless steel
parts with a wet cloth. Customary cleaning agents suitable for stainless steel can be
used without risk. For surface sanitization, use commercially available disinfectants
that are suitable for stainless steel.
Do not use any cleaning agents on the stainless steel housing that contain sodium
hydroxide or acetic, sulfuric or citric acid. The use of steel wool cleaning sponges is
prohibited.
All stainless steel surfaces can be cleaned by wiping with a moist cloth and with the
following products:
– Peracetic acid, H
O
and acetic acid (2.5 %)
2
2
– Quaternary ammonium
– Formaldehyde and glutaraldehyde
– Isopropyl alcohol
– Ethanol 70%
Autoclavable at 121°C for 20 min. are bow, lever, speed selecting knob, support rod,
holding ring for bottles, counter-holding ring for bottles and draining container
(complete).