Fig. 15
20
Configuring the Sterisart Pump
Programs (Fig. 15):
Using this menu item, adjustments can be made to the program step that is currently
activated. In detail, these are:
Speed Minimum
Default setting:
10
Minimum value:
10
Maximum value:
100
This parameter displays the lower limit value of the speed of the current program step
down to which the later user may change the speed.
Standard Speed
Default setting:
Current speed
Minimum value:
Speed Minimum
Maximum value:
Speed Maximum
This value displays the speed of the current program step. The operator can change
the speed within the limit values Speed Minimum and Speed Maximum.
Speed Maximum
Default setting:
100
Minimum value:
10
Maximum value:
100
This parameter displays the upper limit value of the speed of the current program
step up to which the later user may change the speed
Speed Steps (Increments)
Default setting:
10
Minimum value:
1
Maximum value:
10
This value represents the amount the current speed is increased or reduced by turning
the wheel by one position.
Speed Adjustment Timeout [s]
Default setting:
0 s (Speed soft key always with white background)
Minimum value:
0 s (Speed soft key always with white background)
Maximum value:
120 s
The Speed soft key looses the white background (becomes inactive) if the key is not
pressed before this time expires. The soft key would then have to be pressed again
to make a new change to the Speed value.
The value 0 means that the time period does not end, i.e. the soft key no longer
deactivates automatically.
Duration Minimum [s]
Default setting:
0 s (infinite pump time)
Minimum value:
0 s (infinite pump time)
Maximum value:
1800 s
This parameter represents the lower limit value of the current program step's dura-
tion down to which the later user may change the duration