Fig. 16
Duration [s]
Default setting:
Current duration
Minimum value:
Duration mininum
Maximum value:
Duration maximum
This value displays the duration of the current program step. The operator can change
any value between Duration Minimum and Duration Maximum.
Duration Maximum [s]
Default setting:
300 s
Minimum value:
0 s
Maximum value:
1800 s
This parameter displays the upper limit value for the pump time up to which the later
user may change the duration.
Increment Duration
Default setting:
10 s
Minimum value:
1 s
Maximum value.
300 s
This value represents the amount that the current duration will be increased or
reduced by turning the selecting knob around one position.
Duration Adjustment Timeout [s]
Default setting:
0 s (Duration soft key always with white background)
Minimum value:
0 s (Duration soft key always with white background)
Maximum value:
120 s
The Duration soft key looses the white background (becomes inactive) if the key is not
pressed before this time expires. The soft key would then have to be pressed again to
make a new change to the Duration value.
The value 0 means that the time period does not end, i.e. the soft key no longer
deactivates automatically.
Behavior of the foot switch
Standard setting:
Tactile
Minimum:
Switch
Maximum:
Tactile
Use tis setting to adapt the behavior of the foot switch. In the "switch" setting, you
can start the rotor and put it on hold by operating the foot switch. In the „tactile"
setting, the rotor will keep turning as long as the food switch is pressed. Releasing the
foot switch stops the rotor.
Display (Fig. 16)
Contrast Display
Default setting:
50
Minimum value:
10
Maximum value:
100
Use this parameter to set contrast of the display.
Configuring the Sterisart Pump
21