Fig. 9
To read in a barcode, hold it under the laser beam within 10 seconds after you have
pressed the Scan key (see Fig. 9). If no barcode is recognized or read in during this
time, press the Scan key again to repeat this procedure. A successful scanning process
is confirmed by an acoustic signal. At the same time, the barcode recognized is shown
on the display for approx. 1 second.
Follow the same procedure to scan in the 5 additional barcodes in the order given
below:
– Scan sterility test unit
– Scan sample
– Scan rinsing fluid
– Scan tryptic soy (TSB)
– Scan thioglycollate (FTM)
You can press the Esc key at any time during scanning to skip an item.
After scanning, the pump will be in the basic operating state as in Manual Program
(see section 6.1.1)
At the end of the sterility test, the data records collected can be output using a PC
or an optionally available MMC/SD card reader.
For more on the topic of data evaluation, see separate instructions for use.