Location Register read
0xA07E
Mix 1 actual setting
(Read only)
0xA07F
Mix 2 actual setting
(Read only)
0xA080
Mix 3 actual setting
(Read only)
0xA084
Actual Vacuum
buffer pressure
0xA06E
Vacuum buffer
set pressure
(read only)
0xA072
Vacuum buffer
hysteresis
(read only)
0xA076
Vacuum buffer
min on time
(times 100ms)
(read only)
0xA077
Vacuum buffer
min off time
(times 100ms)
(read only)
0xA0D6
Output valve mode 1 byte unsigned
0xA035
"Mixer table 1"
offset
(read only)
Dansensor A/S
DE
Encoding
1 unsigned byte
Range 0 to 100%
For a 3-gas mixer this is the mixer setting of %CO
For a CO
/N
mixer this is the mixer setting of %N
2
2
For a N
/O
mixer this is the mixer setting of %N
2
2
For a CO
/O
mixer this is the mixer setting of %CO
2
2
1 unsigned byte
Range 0 to 100%
For a 3-gas mixer this is the mixer setting of %N
For a CO
/N
mixer this is the mixer setting of %CO
2
2
For a N
/O
mixer this is the mixer setting of %O
2
2
For a CO
/O
mixer this is the mixer setting of %O
2
2
1 unsigned byte
Range 0 to 100%
Only present in a 3-gas mixer.
This is the mixer setting of %O
For a 2-gas mixer this register returns zero.
4 byte float
Actual pressure in the vacuum buffer [bar]
4 byte float
Actual set point for the pressure in the vacuum buffer.
Typical 3.0 or 5.0 bars
4 byte float
Typical 0.3 bar. When buffer pressure reach "Vacuum buffer
set pressure" - "Vacuum buffer hysteresis" the valve is turned on.
When the pressure reaches "Vacuum buffer set pressure" it is
turned off.
1 byte unsigned
Not used, will always return 0
1 byte unsigned
Not used, will always return 0
See also write 0x0066
0 = OFF, mixer is stopped
1 = Running (Auto). Normal operation mode.