Read only commands (R "Read memory (Operational commands")
Location Register read
0xA065
Status register
(Read only)
0xA067
Error register
(Read only)
Dansensor A/S
DE
Encoding
2 bytes unsigned integer
Each bit represent a status condition
0x0001: Always 0 as MMP is already initialized.
0x0002: A system FAULT has occurred
LED on mixer is blinking red at 2Hz (fast blink)
Possible error causes are:
• Flow cannot be maintained (to low input capacity
flow/pressure or backpressure to high)
• Any ERROR listed in User Guide
After FAULT situation, the mixer must be restarted by
either changing the mixture, sending stop/start sequence
using command 0x0066 or external signal
0x0004: Not used in MMP
0x0008: One or more gas input pressure is below limit.
LED on mixer is blinking red at 2Hz (slow blink)
0x0010: Gas inlet pressure is OK.
For mixer type CO
pressure is OK.
For all other mixer types this bit is set when Gas 1inlet
pressure is OK.
0x0020: Gas inlet pressure is OK.
For mixer type CO
inlet pressure is OK.
For all other mixer types this bit is set when Gas 2 inlet
pressure is OK.
0x0040: Gas 3 inlet pressure is OK. If the mixer is configured as a
2-gas mixer this bit is 1 (pressure OK)
0x0080: This bit is 1 when mixer is running and GAS is flowing
through the mixer.
0x0100: Not used in MMP
0x0200: Not used in MMP
0x0400: Not used in MMP
2 bytes unsigned integer
Not used in MMP, this register is always 0x0000 in MMP.
COPYRIGHT ©
Bedienungsanleitung
/N
this bit is set when Gas 2 (N
2
2
/N
this bit is set when Gas 1 (CO
2
2
PBI-300863-B 02/2013
37
) inlet
2
)
2