en
|
Disposal
● To maintain the lifetime of the batteries, charging is only performed within a
temperature range of 5 to 35 °C. Outside this temperature range, the
charging is automatically interrupted and resumes automatically after the
temperature is within the range again.
● The charging time is typically 4 hours.
● A new NiMH power pack reaches its full capacity after three complete
charge/discharge cycles.
● Never store the instrument for extended periods without being connected to
a power source (maximum of 2 months) because the internal buffer battery
will drain.
5.6
Replacing the batteries/rechargeable
batteries
1. Switch off the instrument.
2. Loosen the screw (2.0 mm hexagon socket) on the power pack and remove
the power pack.
3. Replace the alkaline batteries with new ones or the rechargeable NiMH
batteries with charged ones - ensure correct polarity.
4. Completely replace the T4 power pack (with sealed rechargeable batteries,
Order No. 83 18 704/83 22 244).
5. Insert the power pack into the instrument and tighten the screw, the
instrument switches on automatically.
● After replacing the power pack T4, it is recommended that a complete
charging is carried out.
5.7
Cleaning
NOTICE
Risk of damage to the instrument!
Abrasive cleaning implements (brushes etc.), cleaning agents and cleaning
solvents can destroy the dust and water filters.
► Clean the instrument by only using a cloth.
● The instrument does not need any special care.
● If the instrument is very dirty, clean it with a cloth.
24
6
Disposal
This product must not be disposed of as household waste. This is indi-
cated by the adjacent symbol.
You can return this product to Dräger free of charge. For information
please contact the national marketing organizations or Dräger.
Batteries must not be disposed of as household waste. They are there-
fore marked with the adjacent symbol. Dispose of batteries at battery
collection centres as specified by the applicable regulations.
7
Technical data
Extract: see the technical manual for details
Ambient conditions:
during operation and stor-
-20 to +50 °C (-20 to +40 °C with type 180AAHC
age
NiMH single cells and T3 alkaline batteries)
700 to 1300 hPa
10 bis 90 % (short-term up to 95 %) relative
humidity
Protection class
IP 54 for instrument with sensors
Alarm volume
Typically 90 dB (A) in 30 cm distance
Operation time
(with 24 hours' use per day, 1 minute alarm per
day)
Alkali battery
Typically approx. 180 hours
(Order no.
83 20 240/
83 22 239)
1)
The Technical Handbook and Instructions for Use/data sheets for the utilized sensors and the PC
software CC-Vision for Dräger X-am 5100 can be downloaded on the product page of the
X-am 5100 at the following Internet address: www.draeger.com. See also the enclosed Instruc-
tions for Use and data sheets for the sensors used.
Instructions for use
1)
®
|
X-am
5100 Cyanogen