Filter level
Set the amount of signal filtering.
LOW
MEDIUM = normal stabilization time with normal stability.
HIGH
Auto Zero Tracking
Set the automatic zero tracking functionality.
OFF
0.5 Division = display maintains zero up to a drift of 0.5 graduation per second.
1 Division
3 Divisions
Stability
Set the amount the reading can vary while the stability symbol remains on.
0.5 Division = 0.5 graduation
1 Division
2 Divisions
5 Divisions
Auto Tare
Set the automatic Tare status.
OFF = disabled.
ON = enabled.
"Place container on the pan" will be displayed when an Automatic Tare is about to
start.
A Deactivate button is displayed underneath the text. Press this button to deactivate
the Automatic Tare function
Date and Time
Set the date and time.
Change the format (if desired), then enter the current value.
Date: Touch M/D/Y to reposition the Month, Day and Year.
Time: Touch 24 HRS to change the time format to 12HRS.
Press Save to confirm the new value.
Level Assist
Shows how to level the equipment by adjusting the feet.
= faster stabilization time with less stability.
= slower stabilization time with more stability.
= disabled.
= display maintains zero up to a drift of 1 graduation per second.
= display maintains zero up to a drift of 3 graduations per second.
= 1 graduation
= 2 graduations
= 5 graduations
STX