4.5 Density Determination
The Scout can be used to determine an object's density including solid and liquid.
A density determination kit (not included, buy separately) is designed to be used with Ohaus
Scout STX. However, you may use whatever lab apparatus that suits the requirements for
density measurements. The weigh below hook is recommended to use if the solids weight is
more than 200g.
Prepare the balance as shown right
1. Turn off the balance.
2. Remove the balance platform.
3. Insert they up hook and install the beaker
stand (beaker is not supplied).
4. Power on the balance
5. In the upper left portion of the home
screen select Density Determination.
6. Follow below instructions to determine the
object's density.
The DENSITY DETERMINATION Home screen:
4.5.1 Measuring the Density of a Solid Using Water (default)
Press the Item Settings button to open the
Density Determination Settings screen.
Confirm the following Setups are selected:
 Density Type: Solid
 Auxiliary Liquid: Water
 Porous Material: Off
To adjust the water temperature value,
touch the Water Temp button.
A numeric input window appears.
Enter the actual water temperature, then press
Save.