Confirm that the default value displayed (sinker
volume) is correct. To edit the default values,
touch Sinker Volume.
A numeric input window is displayed.
Key in the desired value, then press Save.
The display returns to the previous screen.
Edit the Liquid temperature if required.
To return to the Density Determination home
screen, touch Back.
Press Start to start the process.
Step 1 of 2 – Weigh the Sinker in Air.
Follow the screen instructions, then press
Accept to store the sinker weight ("in air").
Step 2 of 2 – Weigh the Sinker Submerged
in the Test Liquid.
Follow the screen instructions, then press
Accept to store the sinker weight (submerged
in liquid).
Once the necessary weights have been
determined, the density of the Liquid sample is
displayed in g/cm
3
(along with the weight in air,
weight in water) on the Application screen.
Press Start to reset all the weight values and
restart the process.