Herunterladen Diese Seite drucken

OHAUS Scout STX Serie Bedienungsanleitung Seite 21

Vorschau ausblenden Andere Handbücher für Scout STX Serie:

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

STX
4.5.4 Measuring the Density of Porous Material Using Oil
To enable this feature, enter the Item Settings
screen, and select the following:
 Density Type: Solid
 Auxiliary Liquid: Water
 Porous Material: On
To adjust the Water Temperature or Oil
Density values, touch the Water Temp or Oil
Density button.
The balance calculates water density based
on the water temperature value entered (look-
up table).
Measure the actual water temperature using a
precision thermometer.
Numeric input windows appear.
Key in the desired value, then press Save.
To return to the Density Determination home
screen, touch Exit.
Press Start to begin density determination.
Step 1 of 3 – Weigh the Un-Oiled Sample in
Air.
Follow the screen instructions then press
Accept to store the dry sample weight (in air).