5.2 Main Menu
The Scout balance Main menu structure is illustrated below.
5.2.1 Calibration
Scout STX balances offer a choice of two calibration methods: Span Calibration and
Linearity Calibration.
Note: Do not disturb the balance during any calibration.
Calibration sub-menu
Span
Linearity
Calibration
Calibration
5.2.1.1 Span Calibration
Span calibration uses two calibration points, one at zero load and the other at
specified full load (span). For detailed calibration mass information please see the
specification tables in section 7.
Before beginning calibration, make sure the menu Security Switch is unlocked.
With the balance turned ON and no load on the pan, touch Span Calibration to
initiate the procedure. Additional calibration values to be used are shown on the
display. The best accuracy is achieved using the mass closest to the full span value.
Note: To change the span calibration point, touch the alternate weight shown on the
display.
Follow the screen instructions and place the specified calibration weight on the
balance when prompted to do so. When complete, the display shows the Span
calibration status and returns to the current application.
5.2.1.2 Linearity Calibration
Linearity calibration uses three calibration points; zero, mid-capacity and full
capacity.
Before beginning calibration, make sure the menu Security Switch is off.
With no load on the balance, press Linearity Calibration from the calibration sub-
menu and follow the screen instructions.
Place the specified calibration weight on the balance and the calibration process
completes.
The display shows status, then returns to the current application.