Operation
6.4
Testing the sensitivity
3
1. Replace the filter (1) upstream of the sensor.
2. Apply the gas from the test gas cylinder 50 ppm NH
±5 % with the gas supply adapter (2) to the sensor with
min. 0.5 L/min (see list of accessories on page 12).
The green LED at the sensing head must start flashing
within 5 minutes. The sensor signal in the test window is
50 ppm ±10 ppm. The flashing light changes to continuous
after max. 5 minutes.
The sensor is ok.
18
NOTICE
The sensitivity cannot be adjusted.
A multi-meter with max. value storage function can be
connected to the 0 V to 10 V signal cable at the junction box or
to at the control unit. The voltage value read in the measuring
range 0 V to 10 V corresponds to the measured concentration
in the 0 to 100 ppm NH
1
NOTICE
If the test window is no longer reached, the sensor must be
replaced.
7
Operation
1. Switch on the supply voltage at the control unit.
After activation or after an interruption of the supply voltage the
sensor must warm up for a defined period. This is indicated by
the LED flashing green/yellow.
2
The warmed up fault-free operation of the sensing head is
indicated after max. 30 minutes by the green LED being
00433718.eps
illuminated.
(3)
A faulty operating state is indicated by the yellow LED
3
flashing.
range.
3