Mounting
housing or alternatively with a mounting bracket (see list of
accessories) on the wall.
CAUTION
The sensing head must not be mounted near the fresh air
supply, in front of windows or doors, large electrical
consumers (electromagnetic influences), in exhaust air ducts,
behind air washers or behind screens. The sensing head
must not be mounted in inaccessible areas. Protect against
direct exposure to dust or splash water. For wall mounting,
unobstructed gas entry must be ensured.
The sensing head must be mounted vertically with the sensor
pointing down; ±30° at a location with low vibration and, as far
as possible, stable temperatures.