Remove the S-DAT from the interface.  Do not damage the S-DAT!
1 →
Promass, t-mass
8.4
Reassembling the sensor connection housing
Reassemble the sensor connection housing in the reverse order, as →  18 described in
Section 8.1.
Note the following:
8.5
Note on reassembling the Promag
8.5.1
Standard transmitter
NOTICE
Poor measuring performance!
If the grounding switch is in the wrong position, this impacts the measuring performance!
Check that the grounding switch is in the correct position and correct if necessary!
1 →
Version A: If a grounding switch
is provided on the ISEM: set the
grounding switch of the ISEM to
the "right" position (open).
Endress+Hauser
2
Prosonic Flow
2
Version B: If a jumper is
provided on the ISEM: plug in
the jumper for the ISEM
grounding setting on the outside
(open).
EA01278D
21