With
the
RS
232
communication is established by selecting HOST. Data
are sent to an external computer every time the
external computer asks for the values.
The data available are:
High/low speed adjust
Pump/controllers operating condition
Cycle time
Pump life
Pump temperature
Pump current
Power
Pump voltage
Controller output frequency
Cycle number
K1 condition
K2 condition
Life time and cycle # zeroing
Configuration parameter readings
Configuration parameter setting
Note that the new input value are put in effect only at
the next STOP/START of the pump.
If PRINT is selected and a printer is connected on RS
232
line,
an
unidirectional
established and every minute the data are sent to the
printer, even if the pump is not running.
The set of data available are:
Pump speed KRPM
Pump temperatures
Pump current A
Pump power W
K1 condition
K2 condition
The controller is factory-set to HOST.
Confirm the selection by pressing the PUMP
CURRENT push-button.
In this way you enter into an operating phase
named "Monitor Relay Programming" described in
the following paragraph.
connected,
a
bi-directional
communication
In case of RS 485 interface, the controller will display:
Press the PUMP CURRENT push-button to enter
the value and if the serial RS 485 option is installed,
the display shows:
S E R I
is
where: XX = is the RS 485 address range [00-31]
The controller factory set is 00.
Select the desired value by pressing the CYCLE
NUMBER
S E R I
84
TECHNICAL INFORMATION
A L
A D D R E S S
X X
A L
A D D R E S S
0 0