Window-Based Protocol
The following table, valid for the RS 232, RS 422 and RS 485 interfaces, describes each single window used in the
protocol.
WIN
R
W
T
000
X
X
L
001
X
X
L
100
X
X
L
101
X
X
L
102
X
X
L
103
X
X
N
104
X
X
N
106
X
X
N
107
X
X
N
108
X
X
N
109
X
L
200
X
N
201
X
N
202
X
N
203
X
N
205
X
N
206
X
N
207
X
L
208
X
L
209
X
N
210
X
N
300
X
N
301
X
N
302
X
N
400
X
A
402
X
A
WIN = Window
R
= Read
W
= Write
START/STOP
LOW SPEED [0=OFF / 1=ON]
SOFT START [0=NO / 1=YES]
DEAT TIME
[0=NO / 1=YES]
PROCESS GAS [0=Argon / 1=Air] Default Argon
SPEED THRESHOLD
RUN UP TIME (0÷359.999 sec)
HIGH SPEED ADJUST
MODE (0, 1, 2) [FRONT, REMOTE, SERIAL]
BAUD_RATE (0-4) [600, 1200, 2400, 4800, 9600]
PUMP LIFE RESET (TYPE "ON" TO RESET)
CURRENT [ A]
VOLTAGE [ V]
POWER
[ W]
FREQUENCY [ Krpm]
PUMP STATE (0÷6) [STOP, WAITING INTERLOCK, STARTING, NORMAL,
HIGH LOAD, FAILURE, APPROACHING]
ERROR CODE (0÷7) [NO ERROR, OVERVOLTAGE, SHORT CIRCUIT,CHECK CONN.,
TOO HIGH LOAD, OVERRIDE, OVERTEMP PUMP, OVERTEMP CONTR.]
STATE K1 [0=OFF / 1=ON]
STATE K2 [0=OFF / 1=ON]
UPPER BEARING TEMPERATURE (0÷90) [ °C]
LOWER BEARING TEMPERATURE (0÷90) [ °C]
CYCLE TIME
(0-999.999) IN MINUTES
CYCLE NUMBER
(0- 65.535) IN COUNTS
PUMP LIFE
(0-999.999) IN HOURS
CRC PROGRAM LISTING
CRC PARAMETER LISTING
T = Type:
L = Logical
N = Numeric
A = Alphanumeric
DESCRIPTION