Press CYCLE NUMBER to select YES or NO.
S O F T
S T A R T
S E L E C T I
Enter the
selection
CURRENT push-button, and the display shows:
P R O C E S S
X X X X X
where: XXXXXX = Argon or Air
If Argon is selected the maximum power delivered to
the eletrical motor allows the pumps to work properly
with heavy gases such Argon, Silane and Arsine). If Air
is selected the maximum power delivered to the
electrical motor allows the pump to work properly with
light gases such as Nitrogen, Helium and Hydrogen.
The controller is factory set to Argon.
M O D E
O N :
Y E S
by pressing
the
PUMP
G A S
?
Press CYCLE NUMBER to select Argon or Air.
P R O C E S S
A R G O N
Enter the
selection
CURRENT push-button, and the display shows:
F R O N T
S E L E C T
where: XXXXXX = means the word FRONT. REMOTE,
or SER depending on the last selection. The controller
is factory-set for FRONT panel operation.
Choose the desired selection by pressing the
CYCLE push-button.
F R O N T
S E L E C T
82
TECHNICAL INFORMATION
G A S
by pressing
/
R E M O T E
/
I O N : X X X X X X
/
R E M O T E
/
I O N :
F R O N T
87-900-913-01 (N)
?
the
PUMP
S E R
S E R

Werbung

loading

Diese Anleitung auch für:

969-9462969-9562