Use in systems with unstable gases
and liquids such as hydrogen
cyanide, dissolved acetylene or NOx
is not permitted.
Devices must be protected from solar
radiation and rain.
Pressure monitors are precision
devices, which are calibrated in the
factory. Never open the device and
do not change the adjustment of the
varnished calibration screws.
Protect the pressure monitor against
excessive vibrations, e.g., by means
of mechanical isolation or other
vibration damping measures.
Heavily contaminated media can clog
the sensor and cause errors and/or
malfunction. If the equipment is being
used for this purpose, suitable
chemical seals must be connected.
40
Safety Instructions
Pressure monitors and chemical
seals form a functional unit and must
not be disconnected from each other
in the field.
Before disassembly (removing the
pressure monitors from the system),
the device must be disconnected
from the power supply and the
system must be emptied. Observe
the Accident Prevention Regulations.
Never use the pressure monitors as a
climbing aid.
Honeywell GmbH accepts no liability
for non-compliance.