Alternating current measurement in mA
• Connect the black measuring line to the COM socket "8" and the
red one to the socket "DC/AC mA 100 mV".
• Set the adjustment wheel "4" to the area for current measurement
DC/AC and the slider "5" to "AC"
• Connect the two measuring prods with the object to be measured.
Make sure that the given measurement object is disconnected
from the electrical current.
• The measuring value is shown on the analogue scale "1". Take the value and multiply it with
the values of the following table:
AC range (mA)
0,25 mA
2,5 mA
25 mA
50 mA
250 mA
500 mA
Direct current measurement 10 A
This connection is designed for a limited period of currency
measurement only.
• Connect the black measuring line to the COM socket "8" and
the red one to the socket "DC/AC mA 100 mV".
• Set the adjustment wheel "4" to the area for current "10 A"
and the slider "5" to "DC,
• The measurement object must always have no electrical current passing. Observe that the
measurement does not take any more than 15 seconds. Wait at least 30 seconds between
any two measurements!
• The measured value is shown on the analogue scale "1" from 0 – 10
34
Analogue reading scale
250
250
250
50
250
50
, Ω"
DC
V
AC
Multiplication factor
x 0,001
x 0,01
x 0,1
x 1
x 1
x 10
DC
V
AC
DC
AC
DC
AC