USER MANUAL FOR
LOCKING:
With the hoist legs apart (opening pedal on the right hoist leg), lock
the pedal by pushing the locking handle downwards.
UNLOCKING:
Pull the locking handle upwards to unlock the pedal, then close the
hoist legs pressing the left pedal.
VI-2) 2-POINT HANGER BAR
Available as an accessory or as a replacement option for the
standard 4-point hanger bar.
28
SAMSOFT 175
VII - SLING RANGE FOR THE
SAMSOFT 175
Transfer slings for 4-point lifting beam. Maximum load 250 kg.
See the documentation specific to Drive DeVilbiss Healthcare
France slings.
Do not exceed the maximum load of the patient lift.
Bath net.........................................................................................M size
Bath net.......................................................................................... L size
Comfort polyester......................................................................M size
Comfort polyester....................................................................... L size
Bath netted hammock..........................................................ONE size
Toilet netted hammock........................................................ONE size
VII-1) CONTROLS
(To be performed by an authorised person)
Before each use, check the condition of the slings: no loose seams,
the fabric and slings must not be frayed, no tears or traces of wear
(transparent fabric weave).
VII-2) GENERAL CLEANING
(Whenever necessary; to be performed by an authorised person)
● The polyester slings can be washed at 40°C maximum. For other
slings, please refer to the instructions included with the sling.
VIII - MAINTENANCE OF THE
SAMSOFT 175
Unplug the power cable and remove the battery pack before
servicing.
VIII-1) CLEANING
Monthly or as required; to be carried out by competent personnel.
● Clean the patient lift with a soft cloth soaked in a soap solution,
then carefully rinse and dry.
● Do not use too aggressive detergents, do not clean under
running water.
VIII-2) DECONTAMINATION
(Whenever necessary; to be performed by an authorised person).
● Disinfection is done by spraying detergent and disinfectant
solutions on the external surfaces of the patient lift.
It is a foaming solution that does not require rinsing but simply
wiping. Our internal services use ANIOS solutions.
VIII-3) DAILY MAINTENANCE
Charge the battery pack every night.