USER MANUAL FOR
IV-1.2) REMOTE CONTROL
Operating indicator light:
Under normal operating conditions, the indicator light turns
yellow when and while any function is pressed (it briefly
turns red when the key is released, and then goes off).
In the event of a relay fault in the power supply, the light
turns solid red.
IV-1.3) POWER UNIT
1
2
4
3
1- Emergency stop push button
2- Control panel
3- Connection socket
4- Control box
5- Battery pack (removable)
6- Battery pack release button
Emergency stop push-button
In case of emergency, the central supply can be stopped by
pressing the emergency stop button. All the functions of the
electric system are then deactivated.
The emergency stop button is reset by turning
the button slightly clockwise until it pops out.
The emergency stop button must always be
accessible during operation so that it can be
activated quickly in case of an emergency.
24
SAMSOFT 175
5
6
Control panel
The control panel features:
- a battery charge status indicator
- an electric emergency lowering button
- a maintenance indicator
Battery
The battery charge status displays after
charge status
pressing one of the functions of the
indicator
handset for more than 2 seconds. When
releasing the key, the indicator remains on
for 10 seconds.
Battery level indicators:
Green: the batteries are fully charged.
Green and yellow: the batteries are 75%
charged.
Yellow: the batteries are 50% charged.
Yellow and red: batteries almost totally
depleted.
Charge them.
Red: empty batteries.
They must be charged immediately or their
service life will be shortened.
Note: If several repeated beeps are heard
when the patient lift is rising, this is a
warning indicating the batteries are low.
Electric
If the handset cannot lower the patient,
emergency
a separate boom lowering circuit can
lowering
be activated by pressing the electric
button
emergency lowering button.
Maintenance
If the red light is on (under the small
indicator
spanner symbol), the device requires
preventative maintenance, please contact
Drive DeVilbiss Healthcare France
Customer Service.
IV-1.4) CHARGER AND CHARGING
The charger allows the batteries to be recharged when they are
depleted (see previous chapter on the battery charge status) by
two methods:
Direct charge: by connecting the charger directly to the patient
lift without removing the battery pack.
Please note that it is not possible to operate the actuators with
the remote control during direct charging.
Indirect charge: by removing the battery pack from the patient
lift to connect it to the charger (with this method, you do not have
to move the patient lift closer to a mains socket).