USER MANUAL FOR
IV-1.5) BATTERY PACK
The 2 batteries included in the battery pack do not require any
special maintenance. Though, they must be charged regularly
to preserve their service life (minimum charge every 3 months).
With a second battery pack (available as an option on request),
the device can work 24 hours a day.
IV-2) MANUAL FUNCTIONS:
IV-2.1) MANUAL EMERGENCY LOWERING OF
THE LIFT ACTUATOR
The lift actuator has a manual emergency lowering system that
allows the boom to be lowered manually in the event of a power
failure.
This system consists of a red ring at the end of the cylinder rod
that must be lowered to disengage and turned to lower the
boom.
Caution: Remember to return the ring to the up (engaged)
position after the manual emergency lowering is completed.
26
SAMSOFT 175
IV-2.2) MANUAL OPERATION OF THE HOIST LEGS
(Manual delete version of the hoist legs)
Press the spacing pedal to open or close the
base.
As standard, the patient lift is equipped with an automatic locking
system for the hoist legs in the closed position (toggle lock).
If required, an additional system for locking the side rails in
the open position can be integrated as an option (see relevant
chapter).
IV-2.3) HANGER BAR HOOKS
This patient lift is equipped with a hanger bar with non-return
hooks.
Before transferring the patient, check each loop of the sling is
inside the hooks.
IV-3) CONDITIONS FOR USE
WARNING: The SAMSOFT 175 must only be used indoors on a
level, stable and adherent floor; it must be handled by competent
personnel trained in the use of this type of equipment.
When all the connections are made:
● Make sure the actuator(s) respond correctly to the commands
from the remote control.
● Make sure the batteries are charged enough (see colour
code, chap. IV. 1-c).
If these conditions are met, the device is ready for use.
CAUTION: The actuators should only be used
intermittently: 2 minutes on, and then 18 minutes off.
Do not exceed 2 minutes continuous operation.
Any usage frequency exceeding the above cycle may damage
the lift assembly (actuator/control box). This damage is excluded
from the warranty.