Radiated
electromagnetic
field
IEC 61000-4-3
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Conducted
disturbances
induced by RF fields
IEC 61000-4-6
Voltage dips. short
interruptions
and voltage
variations on power
supply input lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 V/m
80 MHz to 2.7 GHz
2 kV for power
supply lines
1 kV for input/output
lines > 3 m
0.5/1 kV differential
mode
0.5/1/2 kV common
mode
3 V
150 kHz to 80 MHz
0% Un for 0.5 cycle
0 % Un for 1 cycle
70 % Un for 25
cycles
0 % Un for 5 s
30 A/m
IEC 60601-1-2
Portable
Test level
communications
should be used no closer to
any part of the R07A including
cables.
Recommended istance:
d = 1,2√P 80 MHz to 800 MHz
d = 2,3√P 800 MHz to 2,5 GHz
Where P is the aximum power
output of the transmitter
in watt (W) based on the
transmitter's producer and d is
the recommended distance in
meters (m).
IEC 60601-1-2
Mains power quality should be
that of a typical commercial or
Test level
hospital environment.
IEC 60601-1-2
Mains power quality should be
that of a typical commercial or
Test level
hospital environment.
IEC 60601-1-2
Portable and mobile RF com-
Test level
munications
should be used no closer to
any part of the R07A including
cables.
Recommended distance d =
1,2√P
IEC 60601-1-2
Mains power quality should be
Test level
that of a typical commercial or
hospital environment. If the
user of the R07A requires con-
tinued operation during power
mains interruptions. it is re-
commended that the R07A be
powered from an uninterrupti-
ble power supply or a battery.
IEC 60601-1-2
Power frequency magnetic
Test level
fields should be at levels cha-
racteristic of a typical location
in a typical commercial or ho-
spital environment.
and
mobile
RF
equipment
equipment
49