Herunterladen Inhalt Inhalt Diese Seite drucken

9-Electromagnetic Compatibility - Resonance R07A Benutzerhandbuch

Inhaltsverzeichnis
Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 76
9. Electromagnetic compatibility
Guidance and manufacturer's declaration - Electromagnetic EMISSION
The R07A audiometer is intended for use in the electromagnetic environment specified below. The
customer or the user of the R07A should assure that it is used in such an environment.
Emission test
RF Emissions
CISPR11
RF Emissions
CISPR11
Harmonic Emissions
IEC/EN 61000--3-2
Voltage fluctuations/flicker
emission
IEC 61000-3-3
The R07A is suitable for use in the specified electromagnetic environment. The purchaser or user of
the The R07A should assure that it is used in an electromagnetic environment as described below:
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
48
Compliance
Group 1
Class B
Complies
Complies
Producer's guide and declaration - Electromagnetic IMMUNITY
IEC 60601-1-2
Test level
8 kV contact
2/4/8/15 kV air
Electromagnetic environment - guidance
The R07A uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference
in nearby electronic equipment.
The R07A is suitable for use in all establishments,
including domestic establishments and those
directly connected to the public low-voltage
power supply network that supplies building
used for domestic purposes.
Compliance
Electromagnetic
Level
Environment
IEC 60601-1-2
Floors should be wood. con-
crete or ceramic tile. If floors
Test level
are covered with synthetic
material. the relative humidity
should be at least 30%.
Inhaltsverzeichnis
loading