LANGUAGE
Press CHANGE to toggle between the languages available.
AUTOTHRS
Press CHANGE to toggle between NO (Auto-threshold test not active), QUICK or NORMAL. If active, on
the top of the main menu will be displayed the possibility to enter on the Auto-threshold test mode.
Press STORE to save the new setting or EXIT. On exiting, the setting will not be updated.
AUTOFREQ
Press CHANGE to enable or disable this function. If active, this function will switch automatically to
the next frequency, based on the related setting, once a value is stored. This function will help the
operator to reduce the testing time.
FACT RESET
This function resets the device to default settings.
Press CHANGE to select YES or NO, press STORE to confirm your selection.
5.3 Frequencies
To access the R07A Frequencies menu press F2 or one of the two knobs.
10 frequencies are available to enable or disable: 125, 250, 500, 750, 1500, 2000, 3000, 4000, 6000 and
8000 Hz.
1000 Hz cannot be disabled.
Once a frequency has been selected by rotating a knob, simply press on "OFF" or "ON" soft key to en-
able or disable the selected frequency.
Pressing the soft key "Exit", the new settings will be stored and the screen will show again the General
menu.
5.4 Calibration
This section is for trained personnel only and allow to perform the annual calibration of the device.
To enter on this section you need a password.
5.5 About
Press F2 soft key or one of the two knobs to access the information in the About section.
44