Technical Data

12.4 Ambient and Storage Conditions

Temperature
Operation, storage, transport
During filter integrity test
Relative humidity (operation, storage, transport)

12.5 Filter Integrity Test Parameters

Permissible gases
Pressure, maximum*
* Filter-specific test parameters, see data sheet

12.6 Parameters for Emptying the Filter Unit with Compressed Air

Pressure, maximum

12.7 Maximum Operating Time Per Filter Unit

Time
Pressure cycles (one pressure cycle = ∆ p >20%)

12.8 Trolley, Valve Actuators and Handle Service Life

Service Life

12.9 Permitted Filter Units

Maxicaps MR
filter unit type FMRxxxxx*
®
* depending on the configuration
44
Maxicaps
MR Operating Instructions
®
Unit
Value
°C
+5 – +40
Constant ambient temperature
%
< 85
Unit
Value
Compressed air,
nitrogen
bar
2.5
Unit
Value
bar
0.5
Unit
Value
h
48
500
Unit
Value
a
8