IMMUNITY TESTS
Magnetic field at
mains frequency
Conducted RF
Disturbances
IEC 61000-4-6
Radiated RF
disturbances
IEC 61000-4-3
96
TEST LEVEL
IEC 60601-1-2
30 A/m 50 and
60 Hz
3 Vrms from
150 kHz to 80 MHz
6 V in ISM bands
10 V/m from
80 MHz to 2.7 GHz
LEVEL OF
ELECTROMAGNETIC
COMPLIANCE
ENVIRONMENT
Magnetic fields at mains
frequency should have levels of
30 A/m 50 and
representative locations typical
60 Hz
for commercial or hospital
environments.
IPortable and mobile RF
communication equipment must
3 Vrms from
not be used closer to any part
150 kHz to
of the actiTENS, including
80 MHz
cables, than the recommended
separation distance, from the
6 V in ISM bands
equation applicable to the
frequency of the emitter.of the
app. Recommended separation
distance d=1.2 √P
d=1.2 √P for a frequency
between 80 MHz and 800 MHz
d=2.3 √P for a frequency
between 800 MHz and 2.7 GHz
where P is the transient
maximum output power of the
emitter in watts (W) according
to the manufacturer of the
emitter and is the recommended
separation distance in metres
(m). The field intensities of
10 V/m from
fixed RF emitters, determined
80 MHz
by an on-site
at 2.7 GHz
investigation, should be lower
than the level of compliance in
each range of frequencies
Disturbances may be produced
near appliances marked with the
following symbol:
electromagnetic
a
.
b
SBM1AL001 – Notice_202202_V10