Herunterladen Inhalt Inhalt Diese Seite drucken

SUBLIMED actiTENS Gebrauchsanleitung Seite 83

Vorschau ausblenden Andere Handbücher für actiTENS:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

P7 2 HZ BURST
Frequency (Hz)
Frequency (Hz)
Technical specifications
Sequence 1
Sequence 1
Mode of action
Intensity (mA)
Intensity (mA)
Burst
Burst
P8 FREQUENCY MODULATION
Technical specifications
Mode of action
  ƒ 1
sequence (duration: 0.25 s): Gate Control
st
Frequency: 100 Hz
Pulse width: 150 µs
  ƒ 2
sequence (duration: 0.25 s): No pulse
nd
  ƒ These two sequences are alternated every 0.25 s.
  ƒ Default duration: 30 min
Sequence 2
Sequence 2
Inhibition of pain signal
Pause
Pause
  ƒ 1
sequence (duration: 7.5 s):
st
Rising frequency: 2 Hz to 80 Hz
Falling pulse width: 200 µs to 100 µs
  ƒ 2
sequence (duration: 7.5 s):
nd
Falling frequency: 80 Hz to 2 Hz
Rising pulse width: 100 µs to 200 µs
  ƒ These two sequences are alternated every 7.5 s.
  ƒ Default duration: 30 min
Combined action:
  ƒ Inhibition of pain signal
  ƒ Stimulation of endorphin secretion for a general
analgesic effect
Frequency (Hz)
Frequency (Hz)
Time
Time
(seconds)
(seconds)
Frequency (Hz)
Frequency (Hz)
Time
Time
(seconds)
(seconds)
Sequence 1
Sequence 1
Sequence 1
Sequence 1
S S
83

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis