Herunterladen Inhalt Inhalt Diese Seite drucken

KaVo Kerr elements e-motion Bedienungsanleitung Seite 53

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 180
Instructions for use elements e-motion Dental Motor
12 Information about electromagnetic compatibility | 12.3 Guidelines and manufacturer's statement - Electromag-
netic immunity
Interference im-
munity tests
Electrostatic discharge
(ESD) according to
IEC 61000-4-2
Fast transient elec-
trical interference /
bursts according to
IEC 61000-4-4
Surges according to
IEC 61000-4-5
Voltage interruptions,
short-term interrup-
tions, and fluctuations
of the supply voltage
according to IEC
61000-4-11
Magnetic field at a
supply frequency
(50/60 Hz) according
to IEC 61000-4-8
IEC 60601 test
levels
± 8 kV contact dis-
charge
± 15 kV atmospheric
discharge
± 2 kV for power lines
± 2 kV for input and
output lines
± 1 kV push-pull
voltage
± 2 kV common mode
voltage
0 % / 0.5 cycles
at 0º to 315º in 45º
increments
0 % / 1 cycle
70 % / 25 cycles
0 % / 250 cycles
30 A/m
12.3 Guidelines and manufacturer's statement -
Electromagnetic immunity
The elements e-motion is intended for use in an environment of the kind spe-
cified below. The customer or user of the elements e-motion should make sure
that the device is used in an environment of the specified type.
Compliance level
± 8 kV contact dis-
charge
± 15 kV atmospheric
discharge
± 2 kV for power lines
and input and output
lines
± 1 kV push-pull
voltage
± 2 kV common mode
voltage
0 % / 0.5 cycles
at 0º to 315º in 45º
increments
0 % / 1 cycle
70 % / 25 cycles
0 % / 250 cycles
30 A/m
54 / 58
Electromagnetic environment
- Guidelines
Floors should be made of wood
or concrete or be fitted with
ceramic tiles. If the floor is fitted
with synthetic material, the relat-
ive humidity must be at least 30
%.
The quality of the supply voltage
should correspond to that of a
typical business or hospital envir-
onment.
The quality of the supply voltage
should correspond to that of a
typical business or hospital envir-
onment.
The quality of the supply voltage
should correspond to that of a
typical business or hospital envir-
onment.
Magnetic fields at the mains fre-
quency should correspond to the
typical values in a business and
hospital environment.

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis