▶ Press the foot control.
ð
The motor starts.
The torque is limited to the set threshold.
Once the limit is reached, the rotation stops, while the torque is being main-
tained.
Once some load is taken off the motor, the rotation continues.
Once 90% of the set torque value are reached, a notifying beep is issued and
the light starts to pulse.
See also:
2 6.4.2 Setting the torque, Page 32
Note
A notice signal is issued once a level of 90% of the set torque is reached.

6.4.4 File database

The elements e-motion possesses an integrated file database. It is necessary to
check if the file data are up-to-date by comparing them to the respective man-
ufacturer information. Speed and torque of the files in the file database corres-
pond to the maximum permissible values of the respective manufacturer.
The file database includes data on the following file systems and manufactur-
ers:
Manufacturer
File system
COLTENE
FKG
KOMET
MICRO-MEGA
Kerr Endodontics
(SybronEndo)
VDW
Instructions for use elements e-motion Dental Motor
6 Operation | 6.4 Operating the elements e-motion
Manufacturer
▪ HyFlex™ EDM
Dentsply
▪ HyFlex™ CM
▪ BioRace
▪ F360
▪ Hero 642®
▪ Revo-S™
▪ K3™
▪ K3XF
▪ Traverse
▪ LightSpeed
▪ TF™
Twisted Files
▪ TF Adaptive
▪ Mtwo®
35 / 58
File system
▪ ProFile®
▪ ProTaper®
Universal
▪ ProTaper Next
▪ ProFile® GT
▪ PathFile™
▪ GT Series X®
▪ ProFile®
Vortex®
▪ Protaper Gold
▪ Vortex Blue
▪ TRUShape

Werbung