Herunterladen Inhalt Inhalt Diese Seite drucken

Disinfection; Manual External Disinfection; Automated External And Internal Disinfection; Packaging - KaVo Kerr elements e-motion Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 180
Instructions for use elements e-motion Dental Motor
8 Reprocessing steps in accordance with ISO 17664 | 8.2 Disinfection
Note
The reprocessing steps for the elements 8:1 handpiece or the INTRA LUX
motor KL 703 LED are described in the corresponding Instructions for use.

8.2 Disinfection

Note
The unit must be disinfected manually only.

8.2.1 Manual external disinfection

▶ Use a soft disposable cloth and an approved disinfectant for disinfection by
wiping down all visible surfaces of the unit, foot control surfaces, and con-
necting cords. Make sure that all surfaces are fully wetted.
▶ Let the disinfectant act for the prescribed time.
▶ Dry the surfaces.
Permissible disinfectants (application range in accordance with the available
manufacturer's instructions for use and national guidelines. Please note material
safety data sheets.) KaVo recommends the following products based on the
compatibility of the materials. The microbiological efficacy must be confirmed
by the disinfectant manufacturer.
▪ FD 322 (Dürr)
▪ Microcide AF Liquid (Schülke & Mayr)
▪ CaviCide (Metrex)
▪ CaviWipes (Kerr)
Note
Follow the instructions for use of the disinfectant.

8.2.2 Automated external and internal disinfection

The following parts of the unit are released for machine-based disinfection:
▪ elements 8:1 handpiece
▪ INTRA LUX Motor KL 703 LED
Note
The reprocessing steps for the elements 8:1 handpiece or the INTRA LUX
motor KL 703 LED are described in the corresponding Instructions for use.

8.3 Packaging

Not applicable.

8.4 Sterilisation

The following parts are released for sterilisation:
▪ elements 8:1 handpiece
▪ INTRA LUX Motor KL 703 LED
46 / 58

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis