Herunterladen Inhalt Inhalt Diese Seite drucken

Connect The Motor Cord; Connect The Power Supply - KaVo Kerr elements e-motion Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 180

5.1.3 Connect the motor cord

▶ Plug the motor cord onto the motor cord connector on the rear of the ele-
ments e-motion and screw it tight.

5.1.4 Connect the power supply

Power supply and cords/hoses are on the floor.
Slipping and tripping.
▶ Power supply and cord/hoses need to be set-up/routed appropriately such
that they are not on the floor.
Property damage from non-approved power supply.
Property damage to the product.
▶ Operate the product with the power supply (Mat. no. 1.013.4796) exclus-
ively.
Note
The connection of the power supply must comply with the country-specific
regulations and requirements for medical devices.
Note
The power supply automatically adjusts to the available mains voltage.
Note
The protective earth conductor is used as functional earthing (FE) rather than
as protective earthing (PE).
Note
Use the power supply in dry environments exclusively. Make sure that the
power supply is protected from the ingress of liquids.
Instructions for use elements e-motion Dental Motor
CAUTION
NOTICE
27 / 58
5 Startup | 5.1 Connector

Werbung

Inhaltsverzeichnis
loading