Herunterladen Inhalt Inhalt Diese Seite drucken

Mounting The Terminal; Panel-Mount Enclosure - Mettler Toledo IND570 Installationshandbuch

Vorschau ausblenden Andere Handbücher für IND570:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 128
2.3.

Mounting the Terminal

The panel-mount enclosure is designed to mount into a cutout of a flat surface such as an
instrument panel or industrial enclosure or door. The harsh enclosure is designed to be placed on a
desktop or can be mounted to a vertical surface with the mounting brackets included in the terminal
packaging. Mount the terminal where viewing is optimal and the terminal keypad is easily
accessible. Observe location and environment considerations as described in Chapter 1,
Introduction.
2.3.1.

Panel-Mount Enclosure

The panel-mount enclosure includes aluminum clamping brackets at the side of the extrusion. Two
Allen-head set screws are used to tighten the brackets against the panel surface. The enclosure will
mount and seal properly on panel thicknesses from 16 GA to 11 GA.
To install the panel-mount enclosure, follow these steps:
1. Loosen and remove the four Allen-head screws that secure the clamping brackets to the side of
2. Remove the two clamping brackets from the housing.
3. Locate the front panel mounting gasket (Figure 2-7) included with the terminal and remove the
30205319
|
02
|
04/2015
the housing (see Figure 2-6). Use the 2mm Allen wrench included with the terminal.
protective paper to expose the adhesive. Adhere the gasket to the rear of the front panel of the
terminal, ensuring that the gasket is flat and spaced evenly on all sides.
METTLER TOLEDO IND570 Weighing Terminal Installation Manual
Figure 2-6: Clamping Bracket
Clamping
bracket