PRODUCT MAINTENANCE
21. PRODUCT MAINTENANCE
The operator should carry out device checks regularly in the intervals defined by
applicable regulations. Test results must be recorded.
The equipment has been designed and manufactured to keep maintenance to a minimum. The
following work must be performed to preserve the proper and reliable operation of the compressor.
Unauthorised interference hazard.
Repair work outside the framework of standard maintenance (see Chapter 21.1)
may only be performed by a qualified technician (an organisation authorized by
the manufacturer) or the manufacturer's customer service.
Standard maintenance work (see Chapter 21.1) may only be performed by the
operator's trained personnel.
Only use manufacturer-approved replacement parts and accessories.
Danger of injury or equipment damage.
Prior to commencing compressor maintenance, it is necessary to:
-
check if it is possible to disconnect the compressor from the appliance in
order to avoid any risk of injury to the person using the appliance or other
material damage;
-
turn off the compressor;
-
disconnect it from the mains (pulling the cord out of the mains socket);
-
vent the compressed air from the air tank.
Venting compressed air poses an injury hazard.
Wear eye protection, i.e. goggles, when venting compressed air from the
compressed air circuit (air tank).
Burn hazard.
When the compressor is running or shortly thereafter, certain portions of the air
pump may be hot - do not touch these components.
Injury hazard.
Allow
the
connecting/disconnecting the compressed air supply!
The removed grounding conductor during service must be connected back to the
original position after completing the service.
• For any maintenance or repair work, the compressor may be removed from the cabinet by
pulling it from the cabinet (on rolling casters as much as the cord between the compressor and
the cabinet allows). (see chap. 21) The required maintenance or repair work may then be
performed.
• If the cord between the compressor and the cabinet becomes disconnected (by pulling the
mains plug, see chap. 12.2), the compressor shall not be functional. In order to bring the
compressor into function again, please follow, see chap. 12.2 or 21.9 or 21.10.
NP-DK50 PLUS, 2V-ADS-A-3_03-2023
equipment
to
cool
down
before
maintenance,
38
PRODUCT MAINTENANCE
service
or
03/2023