Product description
32
*Interface:
NOTICE
Design of the vacuum system
The collection flask (5) catches particles and droplets at the
inlet of the pump which have been suctioned out of the recipi-
ent contrary to the requirements of the pump. The collection
flask is coated (implosion protection) and attached to the vac-
uum system with a flask clamp (3). The solvents separated at
the pump outlet are collected in the collection flask (2), which
is coated (bursting protection).
A flask clamp (3) fixes the collection flask to the separator
adapter (9).
6.3 Pump
The power switch (6) can be used to switch the pump on and
off. The rotary/push knob (3) can be used to start and stop the
pump and to set the pump flow rate.
Translation of Originial Operating Instruction, KNF 325562-325564 02/23
Vacuum system SH820G / SR820G / SH840G / SR840G
Damage to equipment due to use of
unauthorized accessories at the inter-
face
This interface is provided to allow con-
nection of KNF accessories.
à
Upon request, you will be given an
exact description of the interface.
à
Only ever use accessories that are
authorized by KNF (see the chapter
entitled Accessories).