Setup and connection
Connected com-
ponents
Pump discharge 7. When using as a vacuum pump: Safely drain the pump
26
NOTICE
3. Connect the lines to the pneumatic inlet and outlet.
4. Connect the lines to the pneumatic inlet and outlet. Use a
hose with ID 4 mm.
5. Connect the lines to the pneumatic inlet and outlet. Use a
hose with ID 6 mm.
6. Only connect components to the pump that are designed
for the pneumatic data of the pump (see Chapter 4 Tech-
nical data [} 13]).
discharge at the pneumatic outlet of the pump.
8. Lay the line at the pneumatic inlet and the line at the
pneumatic outlet in a downward slope so that no conden-
sate can run into the pump.
NOTICE
9. Plug the plug of the power cable into a properly installed,
grounded socket.
Translation of Original Operating and Installation Instruction, KNF
Diaphragm pump N86.18 / N811.18
Before mounting the inlet filter or the
silencer, unscrew the corresponding
hose connector from the thread in the
pump head.
Hose radii selected too small can affect
the stability of the pump.
121435-121438 02/23