6.2.2 Cleaning the pipette clamp

The pipette clamp assemblies can be replaced, cleaned or autoclaved as described
below (121 °C, 1 bar overpressure for 20 min.)
Aspirating cone
Pipette adapter
Membrane filter
Sealing
You can also use a membrane filter with a pore size of 0.2 μm.
26 · ErgoOne
FAST · Manual
®
Can be wiped using a damp cloth
Can be disinfected with alcohol (ethanol, propanol)
or alcohol-containing disinfectants.
Repeatedly autoclavable
Can be replaced
Can be rinsed with demineralized water
Repeatedly autoclavable
Can be replaced
To be disposed of if contaminated
Cannot be cleaned
Can be autoclaved once
Can be replaced
Can be rinsed with demineralized water
Repeatedly autoclavable
Can be replaced