15 Instant help for troubleshooting

In case of an error in the program process, briefly turn off the balance and disconnect
from power supply. The weighing process must then be restarted from the beginning.
Fault
The weight display does not
glow
The
displayed
weight
permanently changing
The
weighing
result
obviously incorrect
No data can be transferred to
the WIFI interface
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
52
Possible cause
• The balance is not switched on
• The mains supply connection has been interrupted
(mains cable not plugged in/faulty).
• Power supply interrupted.
• The rechargeable battery / the batteries is/ are
inserted incorrectly or empty
• No rechargeable battery / no battery is/ are
inserted
• Draught/air movement
is
• Table/floor vibrations
• The weighing plate is in contact with foreign
bodies or is not correctly positioned.
• Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
• The display of the balance is not at zero
is
• Adjustment is no longer correct
• Great fluctuations in temperature.
• Warm-up time was ignored.
• Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
• The mains signal is not stable or too weak
• Wrong interface configuration
TMPN_A-BAPS-e-2210

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis