Herunterladen Inhalt Inhalt Diese Seite drucken

Instant Help - KERN CPB-N Serie Betriebsanleitung

Vorschau ausblenden Andere Handbücher für CPB-N Serie:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

14 Instant help

In case of an error in the program process, briefly turn off the balance and disconnect
from power supply. The weighing process must then be restarted from the beginning.
Fault
The displayed weight does not glow.
The displayed weight is permanently
changing
The weighing result is obviously
incorrect
CPB-N / CPB-DM-BA-e-1831
Possible cause
• The balance is not switched on.
• The mains supply connection has been
interrupted (mains cable not plugged
in/faulty).
• Power supply interrupted.
• Batteries are inserted incorrectly or empty
• No batteries inserted.
• Draught/air movement
• Table/floor vibrations
• Weighing plate has contact with other
objects.
• Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
• The display of the balance is not at zero
• Adjustment is no longer correct.
• Great fluctuations in temperature.
• Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
51

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis